CLINICAL GUIDANCE | SECTION TWO
57
ECG monitoring equipment
Caution: During stimulation the recipient must be connected to ECG
monitoring equipment with a member of medical staff qualified in
resuscitation present.
Due to the proximity of the auditory brainstem implant electrodes to the tenth
nerve (vagus), there is a small risk that electrical stimulation could disturb
heart rhythm. A standard 3-lead ECG is sufficient to monitor heart activity.
The recipient may be seated or supine for programming, according to comfort.
You must always follow this procedure for the first fitting and for subsequent
programming if previously untested electrodes are being used, or if the
recipient has reported any side effects consistent with a heart rhythm change
(e.g. feeling unwell, light headed, palpitations, etc.).
2.1.1.2 Programming the ABI
Transmitter coil
If the internal magnet has been removed from the Nucleus 24 ABI implant,
ensure that the external transmitter coil is held in position by use of either
a lightly elasticated headband, or an adhesive retainer disk (available from
Cochlear). In each case, you must ascertain the orientation of the implant to be
able to correctly position the transmitter coil.
To position the transmitter coil:
1. Position the coil over the centre of the superior-posterior section of the
implant behind the pinna.
2. If using a retainer disk, expose an area of skin on the scalp directly over the
receiver coil of the implant.
3. Trim away a circular patch of hair (approx. 15 mm in diameter) a few
millimetres above the centre position of the coil.
4. Wipe clean the exposed skin with gauze moistened with 70% isopropyl
(rubbing) alcohol.
5. When the skin is dry, peel off the backing of the retainer disk and press the
disk firmly to the scalp.
6. Attach the retainer disk so it rests just above the centre of the receiver coil
of the implant, since the transmitter coil will not adhere directly to the
centre of the retainer disk (Figure 3).