SECTION TWO | CLINICAL GUIDANCE
56
2.1.1.1 Setting Up the ABI Equipment
Before switch-on
Since tuning the Nucleus 24 ABI is time-consuming, consider scheduling
the recipient for programming over two to three days. The underlying
medical condition might also mean individual sessions need to be short.
Counsel the recipient about the tuning procedure and assess their
expectations.
Since successful programming of the Nucleus 24 ABI relies upon good
recipient feedback, before each test session and each individual test, give
clear instructions to the recipient
Take a CT scan and/or MRI to check the electrode array has not moved
from the lateral recess. You may take this scan any time between two
weeks after surgery and the morning of the switch-on.
Caution: Recipients who do not have the internal magnet removed from
their implant may be subject to additional regional restrictions with Magnetic
Resonance Imaging (MRI). If the magnet requires removal, see the Nucleus 24
ABI Surgeon’s Guide for more information.
Programming equipment
Use Custom Sound software and a Pod (for a Freedom sound processor) or a
Portable Programming System (PPS) (for a SPrint sound processor) to program
the Nucleus 24 ABI.